MAINTENANCE: Checkout Procedure
7. Check the perfusion indication. Change the selector switch to
"Ohmeda" and verify that the perfusion LED changes to yellow.
Remove the cable from the oximeter. The _ indicator should light in front
of the SpO
value on the oximeter and a "CHECK PROBE" message
2
should appear on the Centralscope.
NOTE
If the Accutracker fails any test, return the unit to GE
Marquette Medical Systems Service and Supplies. See
"How to Reach Us" in Chapter 1.
The following displays on the Accutracker when you turn it on (press and
hold the
button while turning the switch on). This display includes
hardware and software version, the Gain Setting for the mic, and the
pressure setting (mmHg).
Gain Setting
Do the following to ensure the functionality of the Accutracker DX
Noninvasive Blood Pressure monitor.
1. Connect the PDM200 or Mercury manometer to the patient cable/cuff
connection.
NOTE
Make sure no external pressure is applied to the
Accutracker when you turn it on.
APEX Telemetry Transmitter
401566-166
Software Version
Hardware Version
Mercury Pressure in mm
3-15