GE Medical Systems
Information Technologies
to a command sequence of the main software. The energy which was delivered to the patient is
measured and the value transfered to the main software.
Block Diagram PCB Analog (2002316-001)
HV Generation
Shock Control
PCB Digital
Block Diagram
The PCB Digital provides the following functions:
•
Control of all subsystems (Powermanagement, Analog, Pacer).
•
Display
The display is connected to the PCB Digital that contains a graphic controller.
•
Keypad
Any pressed keys are decoded by the PCB Digital and the LEDs on the keypad get illuminated
under control of the PCB Digital.
•
Sound generation
The alarm sounds are generated and amplified on the PCB Digital by a sound chip.
Revision F
2004768-001
Connectors
Paddle Connector Code
Paddle Connector HV
PCB Pacer
HV Capacitor
PCB Power Management
PCB Digital
Speaker
Servicing Manual Dash Responder V1.0