Chapter 6 Method of Use
6.5

Puncture

6-12
Wear protective gear.
There is a risk of infection.
Do not insert the puncture needle into the Ultrasonic Endoscope if you
cannot see the forceps outlet on the endoscopic image. Do not perform
puncture if you cannot see the puncture needle on the endoscopic
image. Operate the puncture needle while checking the endoscopic
and ultrasonic images. Do not insert the puncture needle into the
forceps inlet while the bending portion of the Ultrasonic Endoscope
is bent. Do not insert a bent puncture needle into the forceps inlet.
Stop puncturing immediately if you cannot obtain an endoscopic or
ultrasonic image.
It may cause perforation or bleeding.
Pushing the puncture needle forcefully may damage the Ultrasonic
Endoscope. Do not insert the needle with the sheath of the puncture
needle not protruding from the forceps outlet at the distal end. Do not
insert the puncture needle in to the Ultrasonic Endoscope with the
needle protruding from the sheath.
There is a risk of damaging the Ultrasonic Endoscope.
[Note] If the puncture needle is out of the ultrasonography area, it is not displayed on the
ultrasonic image.
[Note] The puncture needle is bent when it is inserted into the Ultrasonic Endoscope.
The bent puncture needle may not be within the ultrasonography area.
Operate the puncture needle while taking into account the degree of bending.
[Note] The puncture needle remains bent if it is inserted or pulled out while the bending
portion of the Ultrasonic Endoscope is still bent. Straighten the bending portion of
the Ultrasonic Endoscope when inserting or pulling out the puncture needle.
[Note] Use the puncture needle according to its operation manual.