Chapter 1 Safety
Chapter 8 Chemical Disinfection
8.1 Methods of Chemical Disinfection
8.4.1 Injecting Sterile Water
8.5 Cleaning and Disinfection Using Cleaning and Disinfection Devices
Chapter 9 Gas Sterilization
Chapter 10 Storage
1-8
Be sure to clean and disinfect or otherwise sterilize all the channels after every case,
regardless of whether used or not. Remove the balloon before cleaning, disinfecting or
sterilizing the Ultrasonic Endoscope. There is a risk of infection.
After immersing in the chemical solution, rinse off remaining chemical solution with sterile
water. Disinfectant might flow into a patient's body.
Use cleaning and disinfection devices appropriate for this Ultrasonic Endoscope. There is a
risk of infection.
Clean, disinfect and sterilize all the channels after every case, regardless of whether used
or not. Remove the balloon before cleaning, disinfecting or sterilizing the Ultrasonic
Endoscope. Applying gas sterilization to wet parts presents incomplete sterilization.
Proceed gas sterilization after vaporizing water out of Ultrasonic Endoscope. There is a
risk of infection.
Proceed aeration after gas sterilization. Gas remaining in Ultrasonic Endoscope after gas
sterilization is harmful to human body.
Do not store Ultrasonic Endoscope in a carrying case. Storage of Ultrasonic Endoscope in
a carrying case may cause infection.