No communications with the XEVO TQD, TQ-XS, TQ-S micro & SQD2 -
WKB7610
SYMPTOMS
• Unable to control MS in MassLynx tune page
• Loss of comms with MS
• "Instrument Not Present" error message in MassLynx pane (bottom right)
ENVIRONMENT
• MassLynx 4.1
• TQD
• Xevo TQD
• Xevo TQ-XS
• Xevo TQ-S micro
• SQD2
CAUSE
Embedded PC does not have communications with the host PC.
FIX or WORKAROUND
1. Close MassLynx and the MS console.
2. Locate the EPC reboot switch on the front of the instrument (small hole just below the power and status LED) or
for TQD check inside front panel door next to electrical and gas connections .
3. Push a piece of peek tubing or a paper clip into the small hole mentioned in step 2. This is where the EPC reboot
switch is located.
4. Wait three minutes for the EPC to run through its boot sequence.
5. Open MassLynx and confirm that a "Not Scanning" message is in the MassLynx pane (bottom right).
6. Open MS Tune page and confirm vacuum readings are "real" and that everything is communicating properly. Note:
If the MS has vented, go to Vacuum > Pump to start the backing pump.
7. Within the MS tune page, confirm that MS API gas can be turned on, the MS into Operate mode and that there are
sensible readbacks.
Waters - The Science of What's Possible.
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB7610_No_communications_with_the_XEVO_TQD
Updated: Fri, 17 Sep 2021 00:04:37 GMT
Powered by
1

Advertisement

loading

Summary of Contents for Waters MassLynx 4.1

  • Page 1 Waters - The Science of What's Possible. No communications with the XEVO TQD, TQ-XS, TQ-S micro & SQD2 - WKB7610 SYMPTOMS • Unable to control MS in MassLynx tune page • Loss of comms with MS • "Instrument Not Present" error message in MassLynx pane (bottom right) ENVIRONMENT •...
  • Page 2 Waters - The Science of What's Possible. ADDITIONAL INFORMATION Alternatively if the above soft reboot does not work you can try this. Recycle PC. Shut all modules on the UPLC stack. Login to PC. Turned power back on to MS (If the system is vented). Or press the EPC reboot switch as described above if system is under vacuum.

This manual is also suitable for:

TqdSqdXevo tqdXevo tq-xsXevo tq-s microSqd2 ... Show all