The common service operations necessary to maintain the Evidence Drying
cabinet for peak performance are listed below.

Routine Maintenance Schedule

Weekly
 Using 70% ethanol, or a suitable disinfectant, surface disinfect the
inside of the cabinet.
 Using an appropriate glass cleaner, or LabSolutions Glass & Surface
Wipes, Labconco part # 1570000; clean the door and the surface of
the UV lamp.
 Operate the cabinet blower, noting if the Filter System Light turns on.
Monthly (or more often as required)
 Using a damp cloth, or LabSolutions Glass & Surface Wipes,
Labconco part # 1570000; clean the exterior surfaces of the cabinet,
particularly the front and top of the cabinet, to remove any
accumulated dust.
 All weekly activities.
 Check the prefilters; replace both if needed.
Annually
 All monthly activities.
 Replace HEPA and granular activated carbon filters.
Product Service 1-800-522-7658
:
:
g
t
h
e
C
a
b
g
t
h
e
C
a
b
i
n
e
t
i
n
e
t
17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Protector 3390000 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Protector 3390001

Table of Contents