Refer to the following table if the cabinet fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.
PROBLEM
Cabinet blower and
lights won't turn on
UV light not working
UV light is dim or
flickering
Product Service 1-800-522-7658
7
:
7
:
h
o
o
t
i
n
g
h
o
o
t
i
n
g
CAUSE
Unit not plugged into
outlet
Circuit breaker(s)
tripped
Door is open
UV Security Lock
not on
Lamp is defective
Lamp wiring is
disconnected
Defective lamp
ballast
Lamp is defective or
is at end of operating
lifetime
Defective lamp
ballast
CORRECTIVE ACTION
Plug the cabinet into appropriate
electrical service.
Check connection to the top of cabinet.
Reset circuit breakers.
Close door – UV light will not work
with the door open.
The key switch for the UV Security
Lock must be turned on.
Replace defective lamp.
Inspect lamp wiring.
Replace lamp ballast.
Replace defective or worn out lamp.
Replace lamp ballast.
23