Chapter 5: Using The Cabinet
Unloading Evidence
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if
appropriate.
Carefully remove objects in the cabinet. Evidence that was considered
biohazardous when loaded should be handled as biohazardous. If your
protocol required the use of evidence prefilters, remove them now and
store them appropriately.
Wipe-Down
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if
appropriate.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
Inspect and clean the prefilters at the bottom of the door, and at the
top of the liner.
Dispose of rubber gloves appropriately, and have lab coat laundered
properly.
Wash hands and arms thoroughly with germicidal soap.
Shutdown
Close the door. If UV disinfection is desired, insert the UV Security
Lock Key, and turn it to the "ON" position. Turn the Fan/UV Light
switch to "UV". The UV light will light.
Turn the UV Light off after an appropriate disinfection time, typically
1-8 hours.
Shutdown
Close the door and push the switch to the center "Off" position.