Drying in the Cabinet
Start-up
Turn off UV light and turn the UV Security switch to "OFF". Remove
the key, and store it in a safe place.
Turn the switch to "Fan".
Check the condition of the prefilters. If your protocol requires the use
of evidence prefilters, ensure a new set of both door and cabinet
prefilters has been properly installed, according to the instructions in
Chapter 6: Maintaining Your Cabinet. Note if the Filter System light
is on.
Wash hands and arms thoroughly with germicidal soap.
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if
appropriate.
Wipe-Down
Open the door to its full open position.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
CAUTION: Do not use bleach on the stainless steel liner. Use
disinfectants compatible with stainless steel.
Loading Evidence
Load the evidence so that air flows around all surfaces of it. Do not
overload the cabinet. Do not place objects directly against the side
walls.
Do not obstruct the intake prefilter (bottom of the door) or the exhaust
air discharge, on top of the cabinet.
Large objects should not be placed close together.
Slowly close the door and use a padlock (not supplied), or secure it
with a chain of custody seal.
Product Service 1-800-522-7658
Chapter 5: Using The Cabinet
15