Dionex Rfc Specifications - Thermo Scientific Dionex RFC Operator's Manual

Table of Contents

Advertisement

Dionex Reagent-Free Controller
A.

Dionex RFC Specifications

Input Voltage Range
Maximum AC Power Draw
Maximum AC Current Draw
Typical AC Current Draw
Output Operating Range
Ripple and Noise
Overload Protection
TTL Connectors
Temperature Range
a. CC=constant current
b. CV=constant voltage
62
90 – 264 Vac, 47 – 63 Hz
45 W
0.75 A
Dionex RFC-10: 0.4 A
Dionex RFC-30: 0.5 A
AES:
SRS:
EG-1/EG-2: CC:
CR-TC:
AES/SRS/EG: CC: less than 0.01 mA rms and
0.020 mA peak-to-peak
CR-TC: CV: less than 200 V rms and 2 mV
peak-to-peak
Output Drift
AES/SRS/EG: CC: less than 0.1% total drift for 8
hours after an initial warm-up of 10 minutes
CR-TC: CV: less than 0.1% total drift for 8 hours
after an initial warm-up of 10 minutes
A continuously acting constant current circuit
protects the power supply for all overloads,
including a direct short on its power output
TTL inputs operate from 0 to 5 VDC level. Inputs
are active low (less than or equal to 0.5 VDC)
Fuses
Two IEC 127 fast-blow fuses rated 3.15 amp
0 – 40 °C (32 – 104 °F); fan cooling is employed
Weight (net)
2.9 kg (6.5 lbs)
a
CC
: 0 – 150 mA, 0 – 50 VDC
CC:
0 – 500 mA, 0 – 10 VDC
0 – 320 mA, 0 – 45 VDC
b
CV
: 24 VDC, 0 – 170 mA
Doc. 031880-04 1/12

Advertisement

Table of Contents
loading