Important information
WARNING: Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the MDPC-
8127, including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Electromagnetic emissions
The MDPC-8127 is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDPC-8127 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
This MDPC-8127 complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MDPC-8127 is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDPC-8127 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
44
R5913216 /03
MDPC-8127
Compliance
Group 1
Class B
Class D
Complies
IEC 60601-1-2 test levels Compliance level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
Electromagnetic environment –
Guidance
The MDPC-8127 uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
The MDPC-8127 is suitable for use
in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Mains power quality
should be that of a typical
commercial or hospital
environment