Inserting A Sensor Chip - GE Biacore X100 Operating Instructions Manual

Table of Contents

Advertisement

5 Operation

5.4 Inserting a sensor chip

5.4

Inserting a sensor chip

Follow the steps below to insert a sensor chip into the instrument.
Step
1
2
3
4
5
6
54
Action
Open the upper front door of the instrument.
Click the Undock chip icon, or select Undock
Chip from the Tools menu. When undocking
is completed, the Dock Chip dialog is displayed
and the sensor chip lamp on the front flashes.
Pull out the chip slide.
If required, remove the old chip.
Insert the chip with the text on the upper side
and the arrow pointing into the instrument.
Push the chip slide all the way in.
Biacore X100 Operating Instructions 28961142 AD