6 Maintenance

6.4 Component maintenance

6.4

Component maintenance

Check for leaks
The pump compartment is placed behind the lower front door. Once a week, check that
there are no liquid or salt deposits at the positions shown in the illustration below.
1
2
3
Part
1
2
3
4
5
6
7
Check system performance
Run System check and pump calibration from the Tools:More Tools menu to check the
peristaltic pump calibration, detector response, sample injection performance and signal
noise. This tool saves the results of the test as a data item and provides a detailed report
that can be priunted. If any test parameters in the report fall outside acceptance limits,
contact your GE representative.
70
Function
Peristaltic pump
Peristaltic pump tube fittings
Degasser tube fittings (only included with Biacore X100 Plus Package)
Syringe pump tube fittings
Syringe pump
Syringe pump barrel (below the plunger)
Syringe pump plunger fitting
4
5
6
7
Biacore X100 Operating Instructions 28961142 AD