Specifications - GE Biacore X100 Operating Instructions Manual

Table of Contents

Advertisement

8 Reference information

8.1 Specifications

8.1
Specifications
General
Parameter
Automation
Molecular weight detection
Required sample volume
Sample/reagent capacity
Analysis time per sample
Detection limit
Typical range for kinetics:
Association rate constant (k
Dissociation rate constant (k
Number of flow cells
Flow cell height
Flow cell volume
Instrument
Parameter
Dimensions w × d × h
(see also
page
Net weight
Supply voltage
Maximum voltage fluctuation
Frequency
84
Space requirements, on
40)
Value
Maximum 15 samples, 24 h unattended operation
Down to 100 Da in various sample environments
Injection volume + 20 to 30 µL (application depen-
dent)
15 sample/reagent vials, volume 1.5 mL
1 water vial, volume 4 mL
Typically 5 to 15 min
Typically 0.1 nM for >10 000 Da analytes.
Typically 1 nM for <10 000 Da analytes.
3
7
-1
):
10
to 10
M
s
a
-5
-1
10
to 0.1 s
):
d
2
Approximately 0.05 mm
Approximately 0.06 µL
Value
600 × 530 × 575 mm
47 kg
Autorange 100 to 240 V~,
protective earthing
± 10% from the nominal voltage
50/60 Hz
Biacore X100 Operating Instructions 28961142 AD
-1
(for protein-protein interactions)