Product Specification - Kane KM940 Manual

Hand-held combustion analyser
Table of Contents

Advertisement

11.

PRODUCT SPECIFICATION

Parameter
Temp Measurement
Flue Temperature with
probe
Inlet Temperature
Pressure
Gas Measurement
Oxygen
Carbon Monoxide
(standard: H compensated)
Carbon Monoxide
(high range)
Nitric Oxide
(standard)
Nitric Oxide
(low range)
Nitrogen Dioxide
Sulphur Dioxide
Calculations
*2
Carbon Dioxide
Losses
Efficiency
Excess Air
Temp (Nett)
CO/CO
ratio
2
Poison Index
Pre-programmed
Fuels
Resolution
1.0ºC/F
0.1ºC/F
0.01 mbar
*1
0.1%
1ppm
0.01%
1ppm
1 ppm
1 ppm
1ppm
0.1%
0.1%
0.1%
0.1%
1.0ºC/F
0.0001
0.01%
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil, Propane,
Butane, Anthracite, Coke, Coal, Kinsale Gas
Accuracy
+2.0ºC +0.3% reading
+1ºC +0.3% reading
+2% of full scale
+0.2%
+20ppm <400ppm
+5%<5000ppm
+10%>5000ppm
+5% reading from 0.1%
to 10%
+5ppm<100ppm
+5%>100ppm
+2ppm<30ppm
+5ppm>30ppm
+5ppm<100ppm
+10ppm<500ppm
+5%>500ppm
+5ppm<100ppm
+5%>100ppm
+0.3% reading
+1.0% reading
+1.0% reading
+0.2%
+2ºC +0.3% reading
+0.0001
+0.01
Range
0-600ºC
32-1112ºF
0-50ºC/32-122ºF
+150mbar to -150mbar
0-21%
0-10,000ppm
0 - 10%
0-5000ppm
0 -100ppm
0 - 1000ppm
0-5000ppm
0-99.9%
0-99.9%
0-99.9%
0-2885.0%
0-600ºC/32-1112ºF
0-0.9999
0-99.99

Advertisement

Table of Contents
loading

Table of Contents