15. PRODUCT SPECIFICATION

INSTRUMENT
Parameter
Temp Measurement
Flue Temperature
Inlet Temperature
Gas Measurement
Oxygen (0
):
2
Carbon Monoxide
(CO):
Carbon Monoxide
(CO):
Nitric Oxide (NO):
Nitrogen Dioxide
(NO
):
2
H
NitrogenOxide (NO)
Sulphur Dioxide (SO
2
Pressure:
Carbon Dioxide (CO
2
Carbon Dioxide
*2
(CO
)
:
2
*2
Efficiency
:
Hydrocarbons (HC):
*1
Using dry gases at STP
*2
Calculated
LEL = Lower Explosive Limit
Resolution
Accuracy
o
0.1
(C/F)
1.O
o
0.1
(C/F)
1.O
0.1%
-0.1% +0.2%
1ppm
+/-20ppm < 400ppm
5% of reading < 2000ppm
+/- 10% of reading
0.01%
>2000ppm
+/- 5% of reading from
1ppm
0.1% to 10%
+/- 5ppm < 100ppm
1ppm
+/-5% of reading>100ppm
+/-5ppm<100ppm
5ppm
+/-5ppm<100ppm
+/- 10ppm < 500ppm
1ppm
+/-5% of reading >
:
0.01mbar/k
500ppm
pa
+/-5% of reading>100ppm
0.1%
+/-0.5% Full scale
):
0.1%
+/-7% of reading +/-0.4%
)
0.1%
+/-0.3%
0.01%
+/-1%
0-5% Methane (LEL)
KM9106 Operators Manual
o
C +/-0.3% of reading
o
C +/-0.3% of reading
Range
0 - 1100°C, 32 - 2140°F
* Use high temperature
probe for gases >
600°C/1112°F
0 - 600°C
0-999°F
0-25%
0 - 10,000ppm
0 - 10%
0 - 5,000ppm
0 - 1,000ppm
0 - 10,000pm
0 - 5,000pm
0 - 150mbar
0 – 20%
0 - Fuel Value
0-100%
+/-5% of reading
53

Advertisement

loading