Product Specification; Unit - Kane 9206 Quintox User Manual

Flue gas analyser
Table of Contents

Advertisement

PRODUCT SPECIFICATION

UNIT

Temp Measurement
Flue Temperature
Inlet Temperature
*1
Gas Measurement
Oxygen (0
):
2
Carbon monoxide (CO):
(standard: H
compensated)
Nitric oxide (NO):
(high range0)
Nitric oxide (NO)
(low range)
Nitrogen dioxide (NO
):
2
Sulphur dioxide (SO2)
(low range):
Sulphur dioxide (SO
)
2
(high range):
Hydrogen sulphide
(H
S):
2
*1
Gas Measurement
Pressure
)*2
Carbon dioxide (CO
2
*2
Efficiency
*1
using dry test gases at STP
*2
calculated
NB: all ppm reading can be displayed in mg/m
Resolution
Range
o
0.1
(C/F)
0-1100°C
32-2140°F
* Use high temperature probe for gases
>600°C/1112°F
o
0.1
(C/F)
0-600°C
0-999°F
Resolution Range
0.01%
0-25%
1ppm
2000
1ppm
1000
1ppm
100
1ppm
100
1ppm
100
1ppm
2000
1ppm
200
Resolution Range
0.01mbar
0-150 mbar
0.1%
0 – Fuel Value
0.1%
0-100%
3
and can be normalised
Overrange
Reading
30%
-
4000
<100ppm
>100ppm <2000ppm
>2000ppm <4000ppm
5000
<100ppm
>100ppm <1000ppm
>1000ppm <5000ppm
300
<100ppm
>100ppm <300ppm
1000
<100ppm
>100ppm <1000ppm
500
<100ppm
>100ppm <500ppm
5000
<100ppm
>100ppm <2000ppm
>2000ppm <5000ppm
300
<100ppm
>100ppm <200ppm
>200ppm <300ppm
Accuracy
o
1.0
C +0.3% of
reading
o
1.0
C +0.3% of
reading
Accuracy
-0.1% +0.2%
+/-5ppm
+/-5% of reading
+/-10% reading
+/-5ppm
+/-5% of reading
+/-10% reading
+/-5ppm
+/-10% of reading
+/-5ppm
+/-10% of reading
+/-5ppm
+/-10% of reading
+/-5ppm
+/-5% of reading
+/-10% reading
+/-5ppm
+/-5% reading
+/-10% of reading
Accuracy
+ 0.5% Full Scale
+ 0.3%
+ 1%
50

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents