General Equipment Specifications; Environmental Conditions - Bio Compression Systems SC-3004-DL Operating Instructions Manual

Sequential circulator
Hide thumbs Also See for SC-3004-DL:
Table of Contents

Advertisement

GENERAL EQUIPMENT SPECIFICATIONS:

DIMENSIONS:
WEIGHT:
INFLATION:
DEFLATION:
CYCLE TIME:
ELECTRICAL:
APPLIED PART:
PROTECTION AGAINST
ELECTRICAL SHOCK
OPERATION MODE:
PROTECTION MODE AGAINST WATER: IP20

ENVIRONMENTAL CONDITIONS:

FOR OPERATION:
AMBIENT TEMPERATURE:
RELATIVE HUMIDITY:
ATMOSPHERIC PRESSURE:
FOR TRANSPORT AND STORAGE:
ATMOSPHERIC PRESSURE:
RELATIVE HUMIDITY:
ATMOSPHERIC PRESSURE:
SC-3004-DL
4.5" H x 11.75 W x 7.75" D
5.5 lbs
52 Seconds
8 Seconds
13 Seconds / Chambers
120 VAC, 60 Hz, 0.5 A
TYPE B
CLASS II
CONTINUOUS OPERATION
+50°F - +100°F (+10°C -+37.8°C)
30% - 75%
700hPa to 1060hPa
-20°F - + 110°F (10°C- +37.8°C)
30% - 75%
700hPa - 1060hPa
6

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SC-3004-DL and is the answer not in the manual?

Questions and answers

Sandy Bethlenfalvy
April 27, 2025

All the pressure is not releasing from the boots when the machine is turned off. What is the problem?

1 comments:
Mr. Anderson
May 14, 2025

If the pressure is not releasing from the SC-3004-DL boots when the machine is turned off, possible causes include:

1. Garment not properly connected to the pump.
2. Damaged, kinked, or twisted garment hose.
3. Damaged garment.

Check these components and contact Bio Compression Systems if the issue persists.

This answer is automatically generated

Table of Contents

Save PDF