5.
TO SET THE PRESSURE OF EACH CHAMBER IN YOUR GARMENT
5.1 With the pump off, press the "Power" On/Off button.
5.2 The pump will display the hours of usage for 5 seconds. While the hour meter is
being displayed (first 5 seconds) Press and hold the "Up" and "Down" button at
the same time and hold them down for 5 more seconds.
5.3 The next set of numbers will show the current pressure setting in the "LED
Readout" and the chamber associated with that pressure in the "Chamber Num-
ber LED".
5.4 Change the pressure setting using the "Up" or "Down" button until you get to
your desired pressure.
5.5 Press the "Mode" button to move to the next chamber and repeat the steps
above.
5.6 After you have set the pressure in all 4 chambers, press the "Power On/Off"
button and your pump will start to operate.
6.
PROGRAMMING YOUR PUMP TO OPERATE IN DIFFERENT MODES
The SC-3004-DL can be run in two different modes - 1 hour shutoff or continu-
"
ous. To set the mode:"
6.1 With the Pump off, press the "Power" On/Off button.
6.2 The pump will display the hours of usage for 5 seconds. While the hour meter is
being displayed (first 5 seconds) Press and hold the "Mode" button 5 more sec-
onds.
6.3 After 5 seconds, you will see "1Hr." in the "LED Readout". This indicates that
you are in the 1 Hour shutoff mode. In this mode, the pump will operate for
one hour and then automatically turn off. The pump comes preset to this mode
so there is no reason to go through these steps if you wish to run your pump in
the 1 Hour Mode. If you are setting the mode, and want to proceed with this
setting, simply press the "Power" On/Off button and your pump will begin its
cycle.
6.4 If you want to operate in a different mode, after "1Hr." shows press the "Mode"
button again and you will see "COn" which indicates you are in continuous
mode. In this mode, the pump will not shut off until you manually turn it off by
pressing the "Power" On/Off button. Once you see "COn", if you want to pro-
ceed with this setting, press the "Power" On/Off button and your pump will
begin its cycle.
11
Need help?
Do you have a question about the SC-3004-DL and is the answer not in the manual?
Questions and answers
All the pressure is not releasing from the boots when the machine is turned off. What is the problem?
If the pressure is not releasing from the SC-3004-DL boots when the machine is turned off, possible causes include:
1. Garment not properly connected to the pump.
2. Damaged, kinked, or twisted garment hose.
3. Damaged garment.
Check these components and contact Bio Compression Systems if the issue persists.
This answer is automatically generated