Section 7
Specifications
Number
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical -
nominal voltage ±10%
Breaker Requirements
Shipping Weight
Number
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical -
nominal voltage ±10%
Breaker Requirements
Shipping Weight
7-4
Revco CxF
ULT1750-10-A
-10°C (14°F) to -40°C (-40°F) in an 18C to 32C * (64.4F to 89.6F) ambient
87.6.0"W x 40.5" H x 33.8" F-B (222.5cm x 102.9cm x 85.9cm)
Add 7.9" (20.1cm) to F-B for wall spacer/handle.
58.8"W x 28.0"H x 18.5" F-B (149.4cm x 71.1cm x 47.0cm)
17.0 cu. ft. (481.4 liters)
Hermetically-sealed compressor
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid; 1.0" (12.5cm) sub-lids
120VAC, 1 PH, 60 Hz, 9.0 FLA
Operating Range: 108-130VAC
20 Amp, Dedicated Circuit,
20 Amp Time Delay Breaker
764 lbs. (347kg)
ULT2050-10-A
-10°C (14°F) to -40°C (-40°F) in an 18C to 32C * (64.4F to 89.6F) ambient
96.0"W x 40.5" H x 33.4" F-B
(243.8cm x 102.9cm x 84.8cm)
Add 7.9" (20.1cm) to F-B for wall spacer/handle.
66.5"W x 28.0"H x 18.5" F-B
(168.9cm x 71.1cm x 47.0cm)
20.0 cu. ft. (566.3 liters)
Hermetically-sealed compressor
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;
1.0" (12.5cm) sub-lids
120VAC, 1 PH, 60 Hz, 10.0 FLA
Operating Range: 108-130VAC
20 Amp, Dedicated Circuit,
20 Amp Time Delay Breaker
776 lbs. (352kg)
ULT1750-10-D
208-230VAC, 1 PH, 60 Hz, 7.0 FLA
Operating Range: 208-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
ULT2050-10-D
208-230VAC, 1 PH, 60 Hz, 6.5 FLA
Operating Range: 208-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
ULT1750-10-V
230VAC, 1 PH, 50 Hz, 7.0 FLA
Operating Range: 208-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
ULT2050-10-V
230VAC, 1 PH, 50 Hz, 6.5 FLA
Operating Range: 208-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
Thermo Scientific