Technical data
Sanitary compatibility
!
Pharma (CoC)
Overfill protection
Telecommunications
External standards and
guidelines
!
96
Overview of permitted process
connections from → ä 15.
Note!
The gap-free connections can be cleaned without residue using the usual cleaning methods.
Many versions of the Levelflex M meet the requirements of 3A-Sanitary Standard No. 74.
Endress+Hauser confirms this by attaching the 3A symbol.
Certificate of Compliance (CoC)
• See ordering information → ä 8, feature 100 "Additional Option:", option"P".
• Materials in Contact with process made of 316L with Δ ferrite < 3%
• Surface roughness Ra < 0,38 μm/15 μin
• Information on ASME BPE Conformity
SIL 2, for 4 to 20 mA output signal (see SD174F/00/en "Functional Safety Manual").
Complies with part 15 of the FCC rules for an unintentional radiator. All probes meet the
requirements for a "Class A Digital Instrument".
In addition to this, all probes in metallic tanks meet the requirements for a
"Class B Digital Instrument".
The European directives and standards applied can be taken from the associated EC Declarations of
Conformity. In addition, the following also applied for Levelflex M:
EN60529
Protection class of housing (IP-code)
Namur - international user association of automation technology in process industries.
• NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NE 43
Standardization of the signal level for the failure information of digital transmitters.
10.1.9
Additional documentation
Note!
This Additional documentation can be found on our product pages on www.endress.com.
• Technical Information (TI424F/00/en)
• Safety Manual "Functional safety manual" (SD174F/00/en)
• Operating Instruction "Description of Instrument Functions" (BA245F/00/en)
Levelflex M FMP43 with FOUNDATION Fieldbus
74 -
74 -
Endress+Hauser
Need help?
Do you have a question about the Fieldbus Levelflex M FMP43 and is the answer not in the manual?
Questions and answers