14. Quality Control

Each institution must establish its own QC policy.
To check the performance of the device, we recommend measuring the temperature with a
thermocouple. The temperature accuracy can be checked with the NANOCOLOR
(REF 919921) or NANOCOLOR
temperature probes are suitable for checking the device in case of a deviation and for adjusting
the device.
Pressing
in Settings
Press the
key to open the test menu. In the test menu all programmed temperatures are
checked. In this menu the calibration of the device is not changed.
Press the
button to open the calibration menu (Figure 18). In the calibration menu, all
programmed temperatures are readjusted using the base calibration stored in the (USB) T‑Set.
After finishing a calibration, the acceptance of the new values must be confirmed with
Pressing the
button in the calibration menu will restore the device calibration to factory
default. With the heating block NANOCOLOR
can be selected in the two menus
Press the
key to start the selected program. The determined temperature deviations are
displayed in different colors depending on the deviation.
Deviation
< 1 °C
1 < x < 2 °C
> 2 °C
A report is stored in the device for both temperature testing and calibration. This report can be
output using the software supplied (see Data Export).
Manual heating block NANOCOLOR
T Set (REF 919917) provided by MACHEREY‑NAGEL. These
®
takes you to test equipment monitoring with (USB) T‑Set.