10 Repair and Disassembly

Tools Required

CAUTION
When replacing the front cover, do not over-torque the screws. Excessive torque may damage the
plastic screw mountings.
You need the following tools:
Screws Used
The following picture shows the range of screws used in the Avalon FM20/30/40/50 fetal monitors:
140
Flat-head screwdriver, head thickness 0.5 mm to fit
transducer screw
Torx-head screwdriver, size T-10, minimum shaft length
80 mm
Small flat-head screwdriver, 2.0-3.0 mm
Long-nosed pliers
Screw Ejot K30x8, T-10
Screw Ejot 3x10, T-10
Screw Ejot 3x8, T-10
Screw Ejot 2.5x8, T-8
Screw M3x12 with washer, T-10
Screw M3x10 with washer, T-10
Screw M3x8 with washer, T-10