7 Testing and Maintenance
Test Block
Test or Inspection
to be Performed
System
Perform the system
(Sys 1-2)
test according to sub
clause 19.201 of
IEC 60601-
1:2005+A1:2012/
EN 60601-
1:2006+AC:2010, if
applicable, after
forming a system.
System
Perform the system
(Sys 3)
test according to sub
clause 19.201 of
IEC 60601-
1:2005+A1:2012/
EN 60601-
1:2006+AC:2010, if
applicable, after
forming a system.
Key: P = Pass, F = Fail, X = test result value to be recorded
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Evaluation
Safety and Functional Test passed
Repair required at a later date, safety and functional test passed
Device must be taken out of operation until repair and passed tests
Device failed and must be taken out of operation.
Notes:
Next Recurrent Test:
Name:____________________________________________________
Date/Signature:_____________________________________________
90
Expected Test Results
Equipment Leakage Current:
Sys1 <= 100 µA (Normal
Condition)
Sys2 <= 300 µA (Single Fault
Condition)
Protective Earth Leakage
Current if medical electrical
system components are
connected to the same Multiple
Portable Socket-Outlet:
Sys3 <= 300 µA
Record the Results (mandatory for
Philips Personnel only)
What to Record
Sys:PSys1/PSys2 or
Sys:FSys1/FSys2
Sys: PSys3 or
Sys: FSys3
Yes