Philips Avalon FM20 Service Manual page 89

Hide thumbs Also See for Avalon FM20:
Table of Contents

Advertisement

Test Block
Test or Inspection
to be Performed
Noninvasive
Perform the
Blood Pressure
Accuracy Test -- see
Performance
"Accuracy Test" on
Tests
page 83.
Performance
Leakage Test -- see
"Leakage Test" on
page 84.
Performance
Linearity Test -- see
"Leakage Test" on
page 84.
Performance Valve
Test -- page
"Leakage Test" on
page 84.
SpO
Perform the SpO
2
Performance
Performance Test --
Test
see "SpO2
Performance Test"
on page 85.
Safety (2)
Perform Safety Test
(2): Sum of
Functional Earth
and Equipment
Leakage Current -
Normal Condition.
Safety (3)
Perform Safety Test
(3): Sum of
Functional Earth
and Equipment
Leakage Current -
Single Fault
Condition (Open
Earth).
Safety (4)
Perform Safety Test
(4): Applied Part
Leakage Current -
Single Fault
Condition, mains on
applied part.
Expected Test Results
X1 = value displayed by monitor
Difference <= 3 mmHg
X2 = leakage test value
X2 < 6 mmHg
X3 = value displayed by monitor
Difference <= 3 mmHg
X4 = value < 10 mmHg
Value should be between 95%
2
and 100%
NC with mains cable:
Maximum leakage current (X1)
≤ 100 µA
SFC with mains cable:
Maximum leakage current (X2)
≤ 100 µA
Maximum leakage current (X1)
≤ 50 µA
7 Testing and Maintenance
Record the Results (mandatory for
Philips Personnel only)
What to Record
PN:P/X1 or
PN:F/X1
PN:P/X2 or
PN:F/X2
PN:P/X3 or
PN:F/X3
PN:P/X4 or
PN:F/X4
No reporting necessary
S(1):P/X1 or
S(1):F/X1
S(2):P/X2 or
S(2):F/X2
S(3):P/X1 or
S(3):F/X1
Actual Results
89