Maternal ECG Supplemental Information as required by IEC 60601-2-27
Response Time of Heart Rate
meter to Change in Heart Rate
Heart Rate Meter Accuracy and
Response to Irregular Rhythm

Noninvasive Blood Pressure

Complies with IEC 80601-2-30:2009 / EN 80601-2-30:2010.
Performance Specifications
Measurement
Systolic
Ranges
Diastolic
Mean
1
Accuracy
Pulse Rate
Range
Accuracy
(average over
noninvasive blood
pressure measurement
cycle)
Measurement Time
Cuff Inflation Time
Initial Cuff Inflation Pressure
Auto Mode Repetition Times
Venipuncture Mode Inflation
Inflation Pressure
Automatic deflation after
*1: Clinical investigation with the auscultatory reference method
The 5th Korotkoff sound (K5) was used to determine the diastolic reference pressures.
M2735A, M2738A HR change from 80—120 bpm Average: 12 seconds
HR change from 80—40 bpm Average: 15 seconds
+
CL Toco
MP,
HR change from 80—120 bpm Average: 10 seconds
CL ECG/IUP
HR change from 80—40 bpm Average: 12 seconds
M2735A, M2738A Ventricular bigeminy 40—60 bpm
Slow alternating ventricular bigeminy 45 bpm
Rapid alternating ventricular bigeminy 163 bpm
Bidirectional systoles 63—73 bpm
+
CL Toco
MP,
Ventricular bigeminy 40—60 bpm
CL ECG/IUP
Slow alternating ventricular bigeminy 30 bpm
Rapid alternating ventricular bigeminy 70—163 bpm
Bidirectional systoles 63—73 bpm
30—270 mmHg (4—36 kPa)
10—245 mmHg (1.5—32 kPa)
20—255 mmHg (2.5—34 kPa)
Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
40—300 bpm
40—100 bpm: ±5 bpm
101—200 bpm: ±5% of reading
201—300 bpm: ±10% of reading
Typical at HR >60 bpm
Auto/manual: 30 seconds (adult)
Maximum time: 180 seconds (adult)
Typical for normal adult cuff: Less than 10 seconds
165 ±15 mmHg
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60, or 120 minutes
20—120 mmHg (3—16 kPa)
170 seconds
12 Specifications
233