Philips Avalon FM20 Service Manual page 32

Hide thumbs Also See for Avalon FM20:
Table of Contents

Advertisement

3 Installation Instructions
Insert the probe covers into the base station. Up to two units of 16 probe covers each can be
5
stored within the base station.
Insert the thermometer into the base station until it clicks into place.
6
7
Insert the other end of the communication cable into the MIB/RS232 (1) port of the monitor.
Configure the MIB/R232 port at the fetal monitor to iTemp.
8
Switch to the Service Mode.
Select the
menu.
In the
In the
NOTE
When disconnecting the thermometer cable from the base station, DO NOT twist the connector. Pull
on the ring of the connector to disconnect the cable.
32
Main Setup
SmartKey to open the
window select
Setup Hardware
window select
Setup Interfaces
Main Setup
window and select
.
Interfaces
for the MIB/R232 port.
iTemp
Hardware
from the