Electrical Inspection

The instrument has undergone extensive testing prior to shipment. Safety testing at installation is not
required (except in situations where devices are interconnected forming a system).
An extensive self-check may be performed. This recommendation does not supersede local
requirements.
All tests are described in the "Testing and Maintenance" on page 55 chapter of this manual.
NOTE
FM20/FM30 with battery option #E25:
After receiving the equipment, the monitor may be in "shipment mode" to prevent that it is
unintentionally powered on during shipment.
In this case, press the On/Standby button for 10 seconds to bring the monitor back into normal
operation mode, or connect the monitor to AC mains power via the external power supply and the
MSL cable.
Claims for Damage
When the equipment is received, if physical damage is evident, or if the monitor does not meet the
specified operational requirements of the patient safety checks, or the extended self-check, notify the
carrier and the nearest Philips sales/support office at once. Philips will arrange for immediate repair or
replacement of the instrument without waiting for the claim settlement by the carrier.

Repackaging for Shipment or Storage

If the instrument is to be shipped to a Philips sales/support office, securely attach a label showing the
name and address of the owner, the instrument model and serial numbers, and the repair required (or
symptoms of the fault). Resend the device only in safe condition and make sure that it is
decontaminated. If available and reusable, the original Philips packaging should be used to provide
adequate protection during transit. If the original Philips packaging is not available or reusable, please
contact the Philips sales/support office who will provide information about adequate packaging
materials and methods.
NOTE
FM20/FM30 with battery option #E25:
Before repackaging for shipment, switch the monitor on and disconnect the AC mains power (unplug
MSL cable). Press the On/Standby button and hold it for approx. 10 seconds until you hear a beeping
sound. Then release the button. The monitor is now in "shipment mode". This prevents that it is
unintentionally powered on during shipment.
To bring the monitor back into normal operation mode, press the On/Standby button again for
10 seconds, or connect the monitor to AC mains power via the external power supply and the MSL
cable. If available and reusable, the original Philips packaging material should be used for repackaging
to protect the On/Standby button from pressure during shipment.
For data protection use the
traces before shipping the monitor. The function
Data Recording
selecting
Coldstart
function (new with release J.3) to erase all stored fetal/maternal
Erase All
window is opened. With prior releases erase the data by entering the Service Mode and
in the
Setup Hardware
menu.
3 Installation Instructions
Erase All
is available as a pop-up key when the
Stored
27