Testing The Patient Module/Toco+ Transducer: Iup Mode - Philips Avalon FM20 Service Manual

Hide thumbs Also See for Avalon FM20:
Table of Contents

Advertisement

Either:
4
attach electrodes to the M1363A adapter cable, and apply the electrodes to the skin (for example
on the wrists).
or
attach the M1363A adapter cable to a patient simulator.
Note
We do not recommend the use of a specific patient simulator. The use of a patient simulator does
not allow checking the specification of the ECG-functionality; it allows only a check of general
function.
Result: You should see MECG values displayed on the maternal display or annotated on the recorder
trace.
If the test results are not as outlined above, repeat the test with another ECG transducer. If this does
not solve the problem:
The MECG adapter cable may be defective. Replace the adapter cable, and repeat the test.
Check all connections.
Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): IUP Mode
To test the IUP functionality of the patient module or the Toco
1
following:
Manometer.
Expansion chamber.
Three lengths of silicone tubing with a 'T' adapter.
Switch on the monitor and the recorder.
2
Connect the patient module or Toco
3
+
transducer to the fetal monitor.
7 Testing and Maintenance
+
transducer, you need the
95