Code
Text message
F045
Error on the current
output
F052
Faulty configuration
F114
Error real time clock
F122
Double address on
the multisensor com-
munication bus
F123
X-ray alarm
F124
Alarm due to in-
creased radiation
F125
Ambient temperature
too high
F126
Error in the trend re-
cording
F141
Communication error
on the multisensor
communication bus
Tab. 2: Error codes and text messages, information on causes as well as corrective measures
Function check
Code
Text message
C029
Simulation
Tab. 3: Error codes and text messages, information on causes as well as corrective measures
Out of specification
Code
Text message
S017
Accuracy outside the speci-
fication
POINTRAC 31 • Foundation Fieldbus
Cause
Error on the current output
Invalid parameter adjustment
Discharge accumulator
Instrument addresse was assigned sever-
al times
External instruments cause radiation
Radiation above the max. adjustment value
Radiation dose too high
Ambient temperature on the housing out-
side the specification
Instrument error
Secondary instrument does not answer
Cause
Simulation active
Cause
Accuracy outside the specification
9 Diagnostics and servicing
Rectification
Check wiring of the current output
Exchanging the electronics
Carry out a reset
Readjust real time clock
Change instrument addresses
Determine reason for X-ray alarm
In case of brief X-ray radiation: Monitor
switching outputs for this time manually
Determine reason for increased radiation
Cool (heat) the instrument or protect it with
isolation material against cold or radia-
tion heat
Call our service
Check Secondary instruments
Rectification
Finish simulation
Wait for the automatic end after 60 mins.
Rectification
Correct adjustment data
43