9 Diagnostics and servicing
Failure
Code
Text message
F008
Error multi sensor
communication
F013
Sensor signals a
fault
F016
Adjustment data ex-
changed
F017
Adjustment span too
small
F025
Invalid linearization
table
F030
Process value out
of limits
F034
EPROM hardware
error
F035
EPROM data error
F036
Faulty program
memory
F037
RAM hardware error
F038
Secondary signals
failure
F040
Hardware error
F041
Photomultiplier error
42
This status message is inactive by default.
Cause
Additional sensors not switched on
EMC influences
No other sensor available
Error on the current input/digital input
No valid measured value
Connected instruments without function
Values of the min. and max. adjustment ex-
changed
The values of the min. and max. adjustment
are too close together
Wrong or empty linearization table (1074,
1075, 1080, 1100, 1106)
Wrong value in the linearization table (1143,
1144)
Process values are not within the adjusted
measuring range
Electronics defective
Error in the internal instrument communi-
cation
Error during software update
Error in RAM
Connection cable to the Secondary instru-
ment interrupted
Instrument not defined as Secondary in-
strument
One of the Secondary instruments signals
a failure
Instrument defective (1092, 1126)
Temperature outside the specification
(1091)
Error in the measured value recording
Rectification
Check wiring between the sensors
Connect the sensors correctly and make
them ready for operation
Check current input
Check connected instruments (Secondary
instrument)
Correct adjustment data
Correct adjustment data
Create linearization table
Correct linearization table
Repeat adjustment
Exchanging the electronics
Carry out a reset
Exchanging the electronics
Repeat software update
Exchanging the electronics
Exchanging the electronics
Check the connection cable to the Second-
ary instrument
Define instrument as Secondary
Check Secondary instruments
Restart instrument
Exchanging the electronics
Cool the instrument or protect it with isola-
tion material against heat/cold
Exchanging the electronics
POINTRAC 31 • Foundation Fieldbus

Advertisement

Table of Contents
loading

Table of Contents