Vega POINTRAC 31 Operating Instructions Manual page 36

Radiometric sensor for level detection
Table of Contents

Advertisement

6 Set up with the display and adjustment module
Menu
Setup
Display
HART mode
Copy instrument settings
36
Menu item
Measurement loop name
Isotope
Application
Adjustment mode
Adjustment - uncovered
Adjustment - covered
Delta I
Background radiation
Temperature unit
Adjustment mode
Adjustment uncovered
Delta I
Mode - Relay
Lock adjustment
Language
Displayed value
Display unit
With this function you can select the mode.
The sensor offers the HART modes standard and multidrop.
The default setting is standard with address 0.
If the measured value is output via the 4 ... 20 mA output, you must
not switch over to HART Multidrop.
The mode 'Standard', with fixed address 0 (factory setting), means
output of the measured value as 8/16 mA signal.
In Multidrop mode, several sensors are communicating on one two-
wire cable via the HART protocol.
In Multidrop mode, up to 15 sensors can be operated on one two-wire
cable. An address between 1 and 15 must be assigned to each sen-
sor.
With this function
Default value
Sensor
Cs-137
Limit level
Single point adjustment
90000 ct/s
9000 ct/s
only with two-point adjustment
90 %
0 ct/s
° C
1-point
900000 ct/s
90 %
Overfill protection
Released
Selected language
Pulse rate
ct/s
POINTRAC 31 • Foundation Fieldbus