Color-Doppler - GE VOLUSON 730Pro Basic User Manual

Hide thumbs Also See for VOLUSON 730Pro:
Table of Contents

Advertisement

Technical Data / Information
23.13

Color-Doppler

CFM Mode:
Color flow imaging is possible with curved array, linear array and phased array probes.
Display modes:
Color coding steps:
Depth range:
Zero line shift:
Inversion of color direction:
Wall Motion Filter:
Smoothing Filter:
Gain control :
Density (color line density):
Ensemble (color shots per line): 7 to 31
Pulse repetition frequency:
CFM Map:
Frequency range:
Balance:
Max. meas. velocity:
Min. meas. velocity:
Display modes:
Scale:
Automatic moving tissue suppression: yes
23-10
2D/CFM (Single, Dual, Quad); 2D+2D/C
Simultaneous: 2D+M/CFM
Simultaneous triplex mode 2D/CFM+/D
8192 color steps
axial: 0 to B scan range
lateral: 0 to B scan range
17 steps (independent from spectral Doppler)
yes
7 steps
12 steps rising time
12 steps falling time
30dB
9steps
100 Hz to 20,5 kHz (18 steps)
8 different color codes for each probe
1 to 15 MHz depending on the probe,
adjustable in 3 steps (low, mid, high)
from 25 to 255 in 41 steps
5.5 m/sec.
0.3 cm/sec.
V-T (velocity + turbulence)
V
(velocity)
V-P (velocity + power)
T
(turbulence)
P-T
(power + turbulence)
kHz, cm/s, m/s
®
Voluson
730Pro - Operation Manual
105831 Rev. 0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents