Maintenance; Maintenance Schedule - SPECTRO SPECTROTEST Original Operating Instructions

Mobile spectrometer for metal analysis
Table of Contents

Advertisement

7

Maintenance

Information
In order to ensure the availability and operational safety of the equipment, we
recommend having the system checked for correct operation and performance at
least once per year by specialized technicians. To find contact addresses of
service organizations authorized by SPECTRO go to www.spectro.com
7.1

Maintenance schedule

The maintenance intervals specified in the following table are guidelines which
should be observed as a minimum requirement. The maintenance intervals may
vary, depending on the use of the instrument.
To ensure the flawless function of the instrument, proceed with the following
maintenance tasks:
After every measurement
Cleaning the electrode
Start of work
Checking the argon supply
Cleaning the adapter
Checking the focal spot
ICALization
As required
Cleaning the windows of the sample probe
and of the adapter
Cleaning the air filter (housing)
Replacing the air filter (sample probe)
Fine dust filter unit
CO
filter unit
2
SPECTROTEST — 05.12.2017
Necessary tasks
Clean the electrode with a small wire brush.
Necessary tasks
Min. pressure in bottle: 10 bar.
See page 32.
Visual inspection
See page 27.
See page 32.
Every month or more often if necessary See
page 34.
Every month or more often if necessary See
page 35.
Replace every six months or more
frequently if necessary (the system will
display a message) See page 36.
Replace every six months or more
frequently if necessary (the system will
display a message) See page 36
Maintenance
— 31