Error Codes - Welch Allyn 300 Series Directions For Use Manual

Vital signs monitor
Hide thumbs Also See for 300 Series:
Table of Contents

Advertisement

Directions for Use

Error codes

Error
code
C01
C02
C03
C04
C05
C06
C10
--
C7
C20
P
C13
E20-E50 Internal failure. Contact customer service.
Description
Blood pressure
NIBP cycle cancelled by operator.
Unable to calibrate. Either the cuff is being moved excessively during the calibration or
the NIBP unit is not working.
Inflation too rapid. Check for hose constriction. Verify patient type setting.
Inflation too slow. Check for hose leak. Verify patient type setting.
Excessive noise or air leak. Check patient condition, cuff placement and connection. Limit
patient motion.
Pressure measurement outside monitor range.
Cuff pressure was too high. Check the patient's condition.
Sensor problem. Check patient condition and sensor position/contact.
Verify SpO
and pulse rate using an alternative method.
2
The sensor is broken, missing, or incompatible. Contact customer service.
Temperature
The probe is broken or missing or the probe well is missing. Contact customer service.
The temperature probe has poor tissue contact.
Battery failure. Use the AC power adapter.
SpO
2
General
Chapter 4 Alarms and alerts
53