Chapter 5 Reviewing patient data
Batch printing
After a patient has been continuously monitored over some period, a clinician can use the
monitor's batch printing capability to print all of the measurements that were stored in
the monitor's memory over the monitoring period.
When the monitor is configured for batch printing, and while it is in normal operating
mode, press
oldest measurement cycle and working forward. The message PRINTING is displayed
during the batch print, and all monitor controls other than
printing continues until one of the following occurs:
In any of these cases, printing stops immediately.
Note
Stream printing
Alternatively, a clinician can choose to print a continuous stream of vital signs, where the
measurement information is printed as soon as the results of each cycle are recorded.
When the monitor is configured for stream printing, and while it is in normal operating
mode, press
of the following occurs:
Printer output
The printer standard report consists of a header, patient information, patient data (table
heading, date, time, range and alarm flags, and measurement readings), an error legend
and a footer. If the monitor receives a print request and no data is available, it prints only
the header and footer.
to print all readings stored in memory (up to 99 cycles), starting with the
All stored data has been printed.
is pressed again.
A battery failure condition.
A nonrecoverable error condition is detected.
If the monitor is configured for Stream printing but printing is disabled, you can
press and hold
for 3 seconds to immediately start a Batch print job. When this
Batch job ends, the monitor is still configured for Stream printing.
to immediately begin stream printing. Stream printing continues until one
is pressed again; the monitor prints an error log (if any errors occurred) and a
footer, and then printing stops.
A nonrecoverable error condition is detected; printing stops immediately.
A battery failure condition is detected. Printing is suspended as long as the
battery failure continues; if the AC power adapter is connected before the monitor
shuts down, printing resumes.