"Oil-Less" Scroll Laboratory Air
1.0 General Information
1.2.1 EN 61000-6-2 (Cont.)
Guidance and manufacturer's declaration - electromagnetic immunity
The TotalAlert Embedded control system is intended for use in the electromagnetic environment specifi ed
below. The customer or the user of the TotalAlert Embedded control system should assure that it is used in such
an environment.
Immunity test
Electrostatic
±6 kV contact
Discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV diff erential
IEC 61000-4-5
mode
±2 kV common mode
Voltage dips, short
<5 % U
Interruptions and
(>95 % dip in U
voltage variations on
for 0,5 cycle
power supply input
<40 % U
lines
(>60 % dip in U
IEC 61000-4-11
for 5 cycles
<70 % U
(>30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz) magnetic
fi eld
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
4107 9006 25.11
IEC 60601
Compliance level
test level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV diff erential mode
±2 kV common mode
<5 % U
T
T
)
(>95 % dip in U
T
for 0,5 cycle
<40 % U
T
)
(>60 % dip in U
T
for 5 cycles
<70 % U
T
)
(>30 % dip in U
T
for 25 cycles
<5 % U
T
T
)
(>95 % dip in U
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If fl oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment
Mains power quality should be that of a typical
)
commercial or hospital environment. If the
T
user of the TotalAlert 2 requires continued
operation during power mains interruptions, it is
T
)
recommended that the TotalAlert 2 be powered
T
from an uninterruptible power supply or battery.
T
)
T
)
T
Power frequency magnetic fi elds should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
1-4