Maintenance - BeaconMedaes 4107 9006 25 Installation, Operation And Maintenance Instructions

Scroll laboratory air compressor system
Table of Contents

Advertisement

"Oil-Less" Scroll Laboratory Air
Appendix A: Desiccant Dryer
A.4 Maintenance
WARNING:
Always notify the appropriate facility
staff before performing any maintenance
or service procedures on the air system.
Compressed air levels may be aff ected
during maintenance or service procedures.
WARNING:
Compressed air can be dangerous unless
safety precautions are observed in the
use of compressed air and compressed
air equipment.
internal air pressure to the atmosphere
before disassembling any subassemblies
or components and before doing any
work on compressed air equipment. To
vent internal air pressure, follow the
maintenance shutdown instructions.
4107 9006 25.11
Completely vent the
A compressed air dryer should give long and
trouble free operation if the recommended
preventative maintenance program is carried out.
The following is a recommended schedule:
1. Quarterly procedure - Clean the auto drain in
the coalescing fi lter. Monitor the backpressure
on the purging tower. If the gauge reads more
than 0 psig (when purging), check the purge
muffl er for blockage and replace if necessary.
2. Annual procedure - Replace all fi lter
cartridges and purge muffl er(s). Check the
automatic drain function in the coalescing
fi lter. Refer to chart located in Section A.4.1 for
correct system size and part numbers. Contact
BeaconMedæs Service at 1-888-4MEDGAS for
parts.
3. Three-year procedure - Change all annual
parts. Change desiccant, check valves, and
purge valve(s). See chart in Section A.4.1 for
correct system size and part numbers.
A.4.1 Maintenance Interval
Service Interval
Every year
Every year
Every year
Every 3 years
Every 3 years
Every 3 years
Every 3 years
As needed
As needed
As needed
Note: All service components are listed as parts
within the service repair kits in Section 5.2
A-6
Description
Pre-fi lter
After-fi lter
Purge muffl er
Desiccant
Check valve
Tower o-ring
Canister o-ring
Switching valve
Tower pressure gauge
Purge Poppit Valve