5.0 Maintenance
5.2.6 2 - 3 Hp Service Intervals
Greasing of the bearings of the compressor element must be done
with special grease and grease gun sold by BeaconMedaes below, and
according to a specifi c procedure. Contact BeaconMedaes for service.
KIT NUMBER
WHERE USED
2903 0139 00
All Systems
2892 6100 20
All Systems
7.5 Hp Units
1630 0311 00
15 Hp Units
Notes:
1. The Grease Pump Kit does not include the Grease Cartridge or the extension nozzle for 7.5 Hp and 15 Hp
units. These items must be ordered separately.
2. 2 - 3 Hp Service intervas: are 5,000 hours OR every two (2) years, whichever comes fi rst, for the orbiting
bearings compressor lubrication and 10,000 hours OR every four (4) years, whichever comes fi rst, for the
pin crank bearings.
3. In high ambient conditions, the bearings must be greased more frequently: for every 9°F increase above
86°F, the maintenance interval should be reduced by 30%.
5.2.7 2 - 3 Hp Only: 10,000 Hours or Every Four (4) Years Compressor Tip Seal Kits
KIT NUMBER
WHERE USED
4107 4000 54
2 - 3 Hp Units
Notes:
1. Service interval is 5,000 hours or every two years, whichever comes fi rst.
2. An authorized BeaconMedæs Service Technician should perform compressor lubrication and tip seals
replacement. Please contact BeaconMedæs Customer Service Department at 1-800-463-3427 prior to the
maintenance interval to schedule the maintenance.
2
OR with Tip Seal Kits used to support all systems
IMPORTANT NOTE:
DESCRIPTION
Grease Pump Kit
Grease Cartridge
Extension Nozzle for grease
pump
1
DESCRIPTION
Kit - Tip seal 2 - 3 Hp
5-5
QTY
CONTENTS
Grease gun
1
Adaptors
1
Grease cartridge
1
Extension nozzle
1
QTY
CONTENTS
Each kit contains:
(2) Sets of element seals
1 per
(1) Dust seal
unit
(1) Round tube backer
seal
(1) Outlet pipe
4107 9006 25.11

Advertisement

Table of Contents
loading

Table of Contents