"Oil-Less" Scroll Laboratory Air
1.0 General Information
1.2.1 EN 61000-6-2 (Cont.)
Guidance and manufacturer's declaration - electromagnetic immunity
The TotalAlert Embedded control system is intended for use in the electromagnetic environment specifi ed
below. The customer or the user of the TotalAlert Embedded control system should assure that it is used in such
an environment.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and
refl ection from structures, objects and people.
a
Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicticted theoretically with
accuracy. To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey
should be considered. If the measured fi eld strength in the location in which the TotalAlert Embedded control system
is used exceeds the applicable RF compliance level above, the TotalAlert Embedded control system should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the TotalAlert Embedded control system.
b
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
IEC 60601
Compliance level
test level
3 Vrms
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the TotalAlert Embedded control system, including
cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1,2
P
d = 1,2
P
d = 2,3
P
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fi xed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
1-5
80 MHz to 800 MHz
800 MHz to 2,5 GHz