"Oil-Less" Scroll Laboratory Air
Appendix C: TotalAlert Embedded Control System
Figure C.31 Setup Menu
C.7.6 Device Setup
This Device Setup page (Figure C.32) is used to
confi gure the medical air system name, location,
facility name, and contact information.
Figure C.32 Device Setup
1. Click Device Setup to access the Device Setup
page.
2. Enter the new device name.
3. Enter the location
4. Enter the facility description.
5. Enter the contact information.
6. Click the Submit button.
C.7.7 Network Setup
This Network Setup page (Figure C.33) is used
to confi gure the facility network information and
e-mail server confi guration.
NOTE: Obtain the IP Address and DNS Name from
the facility's Information Systems department.
Figure C.33 Network Setup
1. Click Network Setup to access the Network
Setup page.
2. Select Enable or Disable for the DHCP Mode.
3. Enter the IP Address for the DNS server
confi guration.
4. Enter the IP Address and DNS name for the
e-mail SMTP server confi guration.
5. Click the Submit button.
C.7.8 Website Setup
This Website Setup page (Figure C.34) is used to
confi gure the automatic refresh interval for the
website. By selecting an auto-refresh interval, the
web page will be refreshed automatically, keeping
all information current during a viewing period.
C-17