"Oil-Less" Scroll Laboratory Air

6.5 Zero Loss Electronic Drain

6.5.1 Isolation of Zero Loss
Electronic Drain
Before servicing the Zero Loss Electronic
Drain, the drain must be isolated in three
locations.
The following valves must be
closed in the following sequence:
1. Receiver drain valve
2. Top knob on manual drain sight glass
3. Valve after the manual drain sight glass
See Figure 6.5 for valve locations.
Electronic
Drain
Figure 6.5 Electronic Drain Isolation Valves
To bring the electronic drain back into
operation, open each of the valves in the
following sequence:
1. Valve after the manual drain sight glass
2. Top knob on manual drain sight glass
3. Receiver drain valve
6.5.2 Recommended Maintenance
Receiver Drain
Valve
Sight Glass
Knob
Visual check and function verifi cation:
Valve Behind
Sight Glass
6-5
Component Maintenance
activity
Condensate
Visual check
Drain
and function
verification.
Integrated
Cleaning and/
fil-ter
or substitu-tion
of the filter.
Condensate
Replace worn
Drain
out com-
ponents,
subject to
wear and tear,
and clean
condensate
drain.
(*) = in the event of applications that are
particularly heavy, dirty, and poorly ventilated,
it should also be foreseen that this frequency is
doubled.
 Inspect the condensate drain for
external damages and leaks.
 Check the operating state of the
condensate drain by means of the
LED displays on the control panel.
 Push the TEST button in order to verify
the operating status of the discharge
and the correct operation of the valve.
Every
Every
day
12
months.
(*)
X
Periodically.
At least once a
month.
X
4107 9006 25.11