The source control system is comprised of four
(4) diff erent printed circuit boards (PCBs) with
interconnecting wiring (RS485 or Ethernet) for
internal communications between the boards.
See Figure C.2.
1. PCB1: Display Board for 5.7" Touch Screen
Display
2. PCB2: Display Board for 3.5" Touch Screen
Display
3. PCB3: Master Source Controller Board
4. PCB4: Pump Controller Board or Dryer
Controller Board
In a standard laboratory air duplex system, the PCB
confi guration consists of the following quantities
and types of boards:
PCB1 – (1) One for the Main 5.7" display
PCB2 – (2) One for each of the Unit 3.5"
displays
PCB3 – (1) Master Controller
PCB4 – (4) One for each compressor (2) and
one for each dryer (2)
In a standard laboratory air quadruplex system,
the PCB confi guration consists of the following
quantities and types of boards:
PCB1 – (1) One for the Main 5.7" display
PCB2 – (4) One for each of the Unit 3.5"
displays
PCB3 – (1) Master Controller
PCB4 – (6) One for each compressor (4) and
one for each dryer (2)
4107 9006 25.11
C.2 PCB1 (5.7" Display Controller)
C.2.1 Basic Software Architecture
The primary purpose of PCB1 is to drive the LCD
display for the 5.7" Master screen.
functions include the following:
1. Communicate to the Master Controller Board
(PCB3) via a RS-485 bus to relay commands
from the touch screen interface and display
messages from the master controller.
2. Interface to the 5.7" Display touch screen to
interpret the user interaction.
3. Drive the 24VDC local alarm horn when
signaled by the master controller via the RS485
bus.
4. Read the input from the motion detector at the
front panel and relay the status to the master
controller so it can distribute the information
to the other pump controllers and then to their
corresponding 3.5" display boards (PCB2).
5. Accept new fi rmware via the Ethernet jack when
connected to a PC confi gured with genuine
BeaconMedaes software for reprogramming.