"Oil-Less" Scroll Laboratory Air
Appendix B: Dew Point Transmitter
B.8 Dew Point Sensor Calibration
The dew point sensor is shipped to you pre-
calibrated, no user calibration is required. Contact
BeaconMedæs to check accuracy if required.
B.9 Dew Point Sensor Replacement
Check the response time of the sensor by removing
it from the sensor chamber and covering the probe
with your hand. The dew point reading should rise
rapidly. If the dew point does not rise or is slow
to respond, it is time to replace the sensor. To
replace, disconnect the power to the instrument,
unplug and remove the sensor from the dew point
sensor chamber, replace sensor and reassemble.
Description
Sensor
4107 6554 19
Cable
4107 6510 89
NOTE: The Dew Point Sensor is included in the
2-Year System Sensor Kits as shown in section
5.2.4. If ordering the 2-Year System Sensor Kits,
there is no need to order the above Sensor as well.