Initial Start-Up - BeaconMedaes 4107 9006 25 Installation, Operation And Maintenance Instructions

Scroll laboratory air compressor system
Table of Contents

Advertisement

"Oil-Less" Scroll Laboratory Air
3.0 System Operation
Proximity Sensor
Power ON Light
Alarm Buzzer
Figure 3.1 Touchscreen Controls

3.2 Initial Start-up

CAUTION: Complete the prestart-up procedure
before continuing with the initial start-up
procedure.
WARNING:
To prevent electrical shock, ensure that
ALL electrical power to the system is OFF,
including the disconnect switches and
Automatic-Manual-Off touch screens on
the control panel. The facility's supply
circuit breaker should also be locked out.
NOTE: DO NOT ADD OIL TO THE COMPRESSOR.
The design of the Scroll compressor is totally oil-
less. It is not necessary to fi ll the crankcases with
oil.
Inside the control panel, make sure that all unit
printed circuit boards are set to the manual
override "Off " position. This is indicated by the
middle position "X" on the three-position sliding
switch as shown in Figure 3.2.
4107 9006 25.11
Main Screen
Unit Screen # 1
Figure 3.2 Unit PCB Override Switch
3-2
Unit Screen # 2
Manual Override
Switch
O - On Manual
X - Off
A - Automatic