Backup Pressure Switch Set Point Adjustments - BeaconMedaes 4107 9006 25 Installation, Operation And Maintenance Instructions

Scroll laboratory air compressor system
Table of Contents

Advertisement

"Oil-Less" Scroll Laboratory Air
Figure 6.6 Kit_001
Exploded view:
POS
Description
01
Tank and level sensor
02
Electronic board
03
Solenoid valve
04
Filter cap
05
Filter cap O-Ring
06
Double mesh metallic filter
07
Valve assembly O-Ring
08
Solenoid valve spring
09
Solenoid valve pilot
10
Diaphragm
11
Valve assembly
12
M6x25 screws
13
M3x10 screws
14
Flow limiter
15
Flow limiter plastic ring
6.6 Backup Pressure Switch
Set Point Adjustments
The backup switch is set at the factory to the
operating point(s) as stated on the wiring
diagram supplied with the unit. It is good
practice to cycle the switch to determine
actual operating points before proceeding
with readjustment.
CAUTION :
ALWAYS
change
gradually.
ALWAYS check switch setting before
making any adjustments.
DO NOT force slotted adjustment screw
when it becomes diffi cult to turn.
ALWAYS isolate the pressure transducer
before making any adjustments to the
backup pressure switch.
6-7
Qty
-
-
-
1
1
1
1
1
1
1
1
4
4
1
1
pressure
setting
4107 9006 25.11