Test Methods

7.0 Test Methods
The Pooltest 25 provides simple test protocols
for the key parameters required to effectively
control swimming pools, hydrotherapy pools, spas
and hot tubs.
The tests available by number are as follows:
Test Test
No. Name
0
Transmittance
1
Chlorine-Free
2
Chlorine-Total
3
Bromine-Total
4
Ozone
5
PHMB
6
pH-Phenol Red
7
Total Alkalinity
8
Calcium Hardness
9
Cyanuric Acid
10
Iron/1
11
Copper Free/5
12
Copper-Total/5
13
Salt/10,000
14
Sulphate
15
Chloride
16
Total Hardness
17
Aluminium
18
Nitrate
19
Ammonium
20
Phosphate/4
21
Manganese HR
22
Chlorine/250
23
Hydro. Perox./100
31
Chlorine-Free/5
32
Monochloramine
33
Dichloramine
34
Calc. Hard. Salt
35
Urea
77
Water Balance
16
Notes
Select wavelength manually
using up/down arrows
Select range from
the System Mode
Select range from
the System Mode
Available directly, also as a
Follow-On Test from Test 11
Available as a Follow-On
Test from Test 31 only
Available as a Follow-On
Test from Test 32 only
Use when testing Calcium
Hardness in a salt pool
Use specifically in China only
Choose Palintest or Langelier
from the System Mode
Test 1 Chlorine-Free/5 or Chlorine-Free/10
Standard Range: 0-5.00mg/l (ppm) Cl
Extended Range: 0-10.0mg/l (ppm) Cl
Colour change: Colourless - Purple/Red
1 Rinse test cuvette with sample leaving
two or three drops in the tube.
2 Add one DPD 1 or DPD XF tablet, crush
tablet and then fill the test tube with
sample to the 10ml mark. Mix to dissolve
tablet fully and ensure no particles remain.
3 Take photometer reading immediately.
The result may drift on standing.
4 Retain test solution if the Total
Chlorine Follow-On Test is required.
Test 2 Chlorine-Total/5 or Chlorine-Total/10
Standard Range: 0-5.00mg/l (ppm) Cl
Extended Range: 0-10.0mg/l (ppm) Cl
Colour change: Purple/Red from Test 1
increases in intensity
Carry out this test on the solution
remaining from the Free-Chlorine/5 test.
1 If any shock treatment chemicals have
been added to the pool, add one DPD
Oxystop tablet, crush and mix to dissolve.
Stand for one minute before proceeding.
This will prevent a false response caused
by the shock treatment chemicals.
2 Add one DPD 3 or DPD XT
tablet, crush and mix to dissolve.
3 Stand for two minutes
to allow full colour development.
4 Take photometer reading
after two minutes have elapsed.
Note: To obtain Combined Chlorine
residual subtract Free Chlorine result
from Total Chlorine result:
Combined Chlorine =
Total Chlorine - Free Chlorine
(use DPD 1)
2
(use DPD XF)
2
(use DPD 3)
2
(use DPD XT)
2