USB Connection Issues
I cannot download my data
Check the USB mode is set to Hard Disk, not COM
port mode. In Hard Disk mode the data can be
'dragged and dropped' as with a conventional
memory stick and is available in CSV format.
Opening CSV data files can be accomplished by a
number of text editing or spreadsheet programs.
Where do I find the COM port drivers?
The latest drivers are available at
www.palintest.com
COM port drivers are provided for Windows
operating systems (Windows Vista, XP and 7).
Care and Maintenance
The Pooltest 25 contains no user-serviceable
parts internally. User maintenance
is only recommended for cleaning of the optical
chamber, changing batteries and validating
performance using the Check Standard Mode.
Cleaning the Optical Chamber
The optical chamber has been designed to
support removal and cleaning with a lint-free
cloth as required by removal of the access cover.
Do not use any of the following agents
when cleaning this optical chamber:
• Abrasive cloths
• Corrosive chemicals
• Any organic solvents
Do not overtighten the screws on re-assembly
to avoid damaging the access cover.
Appendix 2 - Troubleshooting
Replacing the Batteries
Remove the four retaining screws from the
battery cover and gently prise the cover free.
Replace all batteries at the same time.
Ensure on replacing the battery cover that
the gasket is correctly located to prevent any
water ingress. Tighten the screws carefully
but do not overtighten.
Access
Cover
Screws
10
Battery
Cover
Access
Screws
25