Table 11 Setpoint Voltages At A3Tp11 And A3Tp15 - Short Body Tube - Agilent Technologies 5517B User Manual

Laser head
Hide thumbs Also See for 5517B:
Table of Contents

Advertisement

Table 11 Setpoint voltages at A3TP11 and A3TP15 – short body tube*
Actual
R14 Adjusted
Reading
Voltage
V (TP11)
V (TP15)
0.201
0.281
0.202
0.283
0.203
0.284
0.204
0.286
0.205
0.287
0.206
0.288
0.207
0.290
0.208
0.291
0.209
0.293
0.210
0.294
0.211
0.295
0.212
0.297
0.213
0.298
0.214
0.300
0.215
0.301
0.216
0.302
0.217
0.304
0.218
0.305
0.219
0.307
0.220
0.308
0.221
0.309
0.222
0.311
0.223
0.312
0.224
0.314
0.225
0.315
0.226
0.316
0.227
0.318
0.228
0.319
0.229
0.321
0.230
0.322
0.231
0.323
0.232
0.325
0.233
0.326
0.234
0.328
0.235
0.329
* The setpoint voltages at A3TP11 and A3TP15 listed in this table apply to the 05517-60255 or -60256 short body tube with the 05517-60031
A3 Board, and the 05517-60253 or -60254 short body tube with the 05517-60032 A3 Board.
Actual
R14 Adjusted
Reading
Voltage
V (TP11)
V (TP15)
0.236
0.330
0.237
0.332
0.238
0.333
0.239
0.335
0.240
0.336
0.241
0.337
0.242
0.339
0.243
0.340
0.244
0.342
0.245
0.343
0.246
0.344
0.247
0.346
0.248
0.347
0.249
0.349
0.250
0.350
0.251
0.351
0.252
0.353
0.253
0.354
0.254
0.356
0.255
0.357
0.256
0.358
0.257
0.360
0.258
0.361
0.259
0.363
0.260
0.364
0.261
0.365
0.262
0.367
0.263
0.368
0.264
0.370
0.265
0.371
0.266
0.372
0.267
0.374
0.268
0.375
0.269
0.377
0.270
0.378
Adjustments
Actual
R14 Adjusted
Reading
Voltage
V (TP11)
V (TP15)
0.271
0.379
0.272
0.381
0.273
0.382
0.274
0.384
0.275
0.385
0.276
0.386
0.277
0.388
0.278
0.389
0.279
0.391
0.280
0.392
0.281
0.393
0.282
0.395
0.283
0.396
0.284
0.398
0.285
0.399
0.286
0.400
0.287
0.402
0.288
0.403
0.289
0.405
0.290
0.406
0.291
0.407
0.292
0.409
0.293
0.410
0.294
0.412
0.295
0.413
0.296
0.414
0.297
0.416
0.298
0.417
0.299
0.419
0.300
0.420
0.301
0.421
0.302
0.423
0.303
0.424
0.304
0.426
0.305
0.427
5

Advertisement

Table of Contents
loading

This manual is also suitable for: