Waters TQ Detector Operator's Manual page 131

Hide thumbs Also See for TQ Detector:
Table of Contents

Advertisement

To remove the existing capillary
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The probe and source can be hot. To avoid burn injuries, take
Warning:
great care while performing this procedure.
1.
Remove the probe from the source (see
probe" on page
2.
Use the 2.5-mm Allen wrench to remove the 4 probe end-cover retaining
screws.
The probe and source components can be contaminated
4-7).
Replacing the IonSABRE APCI probe sample capillary
"Removing the IonSABRE APCI
End-cover retaining
screws
5-63